Programs:
2 EMG TEST programs /1 ETS program/3 EMG Game and 5 Scene Training programs.
8 Pre-set stim programs and 1 Custom mode program for Pelvic floor rehabilitation.
14 Custom mode for pain and Muscular motor dysfunction, and postpartum rehabilitation.
Electrical stimulation performance
1) Output parameters: under 500Ω load, the maximum effective value of the output current is ≤50mA, and the step is <1mA. .
2) Pulse width: the range is 50-450μs, the accuracy is ±10μs, and the step is 5μs.
3) Pulse frequency: the range is 1-120Hz, the accuracy is ±10%, and the step is 1Hz.
4) Treatment time: adjustable from 1 to 60 minutes, with a step length of 1 minute and a tolerance of ±3%.
Biofeedback performance
1) Accuracy of the feedback threshold: the error should not be greater than ±10% of the nominal value when measured at the 200Hz frequency point.
2) Indication accuracy: the error is not greater than ±10% or ±2μV, whichever is greater.
3) Resolution (measurement sensitivity): ≤2μV.
4) System noise: ≤1μV.
5) Pass-band: The pass-band of the device should not be narrower than 20Hz-500Hz (-3dB) (not including the notch band).
6) Differential mode input impedance: >5MΩ.
7) Common mode rejection ratio: >100dB.
8) Power frequency notch filter: The device should have a 50 Hz notch filter, and the amplitude after attenuation should not be greater than 5μV (peak-valley value).
9) Measuring range: 0.2-2000μV, the error is not greater than ±10% or ±2μV, whichever is larger.